YSI EXO2 Sonde

Summary



TypeCombination
VendorNot Applicable
ModelYSI EXO multiparameter water quality sondes
S/N23A104835
ID Prefix
CORIOLIX IDysiexo104835
ClassFlowthrough
LocationMain Lab
Conditiongood
Configuration
Native Sample Rate1.0 Hz
UDP Port(s)30313 None
EnabledTrue
 


Image Not Available
 
 

General Description



Multiparameter Sonde measures pH, DO, temperature, salinity, depth, turbidity and total algae (chl-a)

 
 

Data Collection Specs

Native Data Feed
Signal Type Digital
Communication Type Ethernet
Serial Details
Baud Rate 9600
Parity N
Stop Bits 1
Byte Size 8
Timeout (s) 1.0
Polled Start Commands
Polled Serial Commands
Ethernet Details
MAC Address
IP Address
Protocol udp
Source File Details
Path
Filename
Native Data Format
Native Data Description Measurements are provided from the sensor in units for each parameter, calculated from the sensor's on-board calibration.
Native Data Type text
Start of Message Characters
End of Message Characters \r
Sample Data Message 9.104 49.445 32.08 7.84 0.00 1.584 5.80 012825 161500 0.09 89.74 8.42 0.49 2.28 1.17 13.43 # 9.104 49.445 32.08 7.84 0.00 1.584 5.80 012825 161500 0.09 89.74 8.42 0.49 2.28 1.17 13.43 8.999 46.660 30.07 7.93 0.00 3.111 5.74 021725 124500 0.36 105.11 10.01 1.89 1.91 1.18 13.45
Message Parser Format ['^\W?\s*(?P<temp>\d+\.\d+)\s+(?P<conductivity>\d+\.\d+)\s+(?P<salinity>\d+\.\d+)\s+(?P<pH>\d+\.\d+)\s+(?P<pressure>\d+\.\d+)\s+(?P<depth>\d+\.\d+)\s+(?P<battery_voltage>\d+\.\d+)\s+(?P<date>\d{6})\s+(?P<time>\d{6})\s+(?P<chlorophyll>\d+\.\d+)\s+(?P<odo_sat>\d+\.\d+)\s+(?P<odo_mg>\d+\.\d+)\s+(?P<pe>\d+\.\d+)\s+(?P<turbidity>\d+\.\d+)\s+(?P<wiper_voltage>\d+\.\d+)\s+(?P<ext_power_voltage>\d+\.\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 1.0 Hz
Shipboard Network Packet Rate None Hz
Ship to Shore Rate None Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address
Port 30313
Packet Size None
Transformed Data Ethernet Distribution
IP Address
Port None
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source internet_time_server
 

Parameters


Quantity Batt (V) Cable (V) Chl-a (RFU) Depth (m) DO (mg/l) DO Sat (%) P (psi a) PE (RFU) pH ( ) S ( ) SpC (mS/cm) T (degrees_C) Turbidity (FNU)
Plot Color
 
 
 
 
 
 
 
 
 
 
 
 
 
Parameter Full Name Battery power Cable Power Chlorophyll-a Depth Dissolved Oxygen Dissolved Oxygen Saturation Pressure Phycoerythrin pH Salinity Specific Conductance Temperature Turbidity
Parameter Short Name Batt Cable Chl-a Depth DO DO Sat P PE pH S SpC T Turbidity
Parameter Standard Name VOLTBATT VOLTELEC FCNTRW01 ADEPZZ01 DOXYZZ01 OXYSSE01 PREXPR01 FVLTPELS PHXXPR01 SSALPR01 CNDCPR01 PSSTTS01 TURBXXXX
SAMOS Data Category Water Temperature
Description Battery voltage for onboard instrument battery pack Power supplied to EXO Sonde through sea cable via shore power Chlorophyll-a fluorescence Ex/Em 470/685nm Depth as measured by stainless steel strain gauge Optical dissolved oxygen Optical dissolved oxygen percent saturation Pressure in pounds per square in gauge Phycoerythrin fluorescence Ex/Em 525/685nm pH Salinity in parts per thousand Specific conductance of sea water Temperature in degrees Celsius Turbidity
Processing Status raw raw raw internally derived internally derived internally derived raw raw internally derived internally derived internally derived raw raw
Processing Symbol battery_voltage ext_power_voltage chlorophyll depth odo_mg odo_sat pressure pe pH salinity conductivity temp turbidity
Diagnostic True True False False False False False False False False False False False
IOOS Category Other Other Biology Pressure Dissolved O2 Dissolved O2 Pressure Biology Other Salinity Salinity Temperature Other
Units V V RFU m mg/l % psi a RFU mS/cm degrees_C FNU
Data Type float float double double double double double double float double double double double
Gross Minimum Value 9.0 V 11.0 V 0.0 RFU 0.0 m 0.0 mg/l 0.0 % 0.0 psi a 0.0 RFU 0.0 0.0 0.0 mS/cm -5.0 degrees_C 0.0 FNU
Gross Maximum Value 16.0 V 13.0 V 100.0 RFU 33.0 m 20.0 mg/l 500.0 % 10.0 psi a 100.0 RFU 14.0 40.0 200.0 mS/cm 50.0 degrees_C 4000.0 FNU
Recommended Minimum Value 9.0 V 11.0 V 0.0 RFU 0.0 m 0.0 mg/l 0.0 % 0.0 psi a 0.0 RFU 4.0 None None mS/cm -5.0 degrees_C 0.0 FNU
Recommended Maximum Value 16.0 V 13.0 V 100.0 RFU 33.0 m 20.0 mg/l 200.0 % 10.0 psi a 100.0 RFU 10.0 None None mS/cm 35.0 degrees_C 999.0 FNU
Global Minimum Value None V None V None RFU None m None mg/l None % None psi a None RFU None None None mS/cm None degrees_C 0.0 FNU
Global Maximum Value None V None V None RFU None m None mg/l None % None psi a None RFU None None None mS/cm None degrees_C 4000.0 FNU
Local Minimum Value None V None V None RFU 0.5 m 1.0 mg/l 0.0 % 0.0 psi a None RFU 7.0 0.0 None mS/cm -2.0 degrees_C 0.0 FNU
Local Maximum Value None V None V None RFU 7.0 m 20.0 mg/l 200.0 % 7.0 psi a None RFU 8.5 35.0 None mS/cm 20.0 degrees_C 4000.0 FNU
Lower Detection Limit None V None V None RFU None m 0.0 mg/l None % None psi a None RFU None None None mS/cm None degrees_C None FNU
Saturated Value None V None V None RFU None m None mg/l None % None psi a None RFU None None None mS/cm None degrees_C None FNU
Bad Data Value None V None V None RFU None m None mg/l None % None psi a None RFU None None None mS/cm None degrees_C None FNU
Resolution None V None V 0.01 RFU 0.001 m None mg/l 0.1 % 0.001 psi a 0.01 RFU 0.01 None None mS/cm 0.001 degrees_C None FNU
Accuracy +/- None V +/- None V +/- None RFU +/- 0.004 m +/- None mg/l +/- 1.0 % +/- 0.004 psi a +/- None RFU +/- 0.1 +/- None +/- None mS/cm +/- 0.01 degrees_C +/- None FNU
Precision None V None V None RFU None m 0.1 mg/l None % None psi a None RFU None None None mS/cm None degrees_C 0.01 FNU
Binning Type linear linear linear linear linear linear linear linear linear linear linear linear linear
Online Display Format 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f 4.1f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power 12.0 - None V AC
Ideal Power 13.0 V AC
Operating Current None - None mA @ None V AC
Analog Output None - None
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth None - None m
Operating Temperature None - None degree_C
Lamp Life None days
Response Time None seconds
Additional Time Offset None seconds
Physical Status
Current Location installed
Current Status testing in progress
Physical Condition good
Physical Configuration
Software Version
Software Configuration
Data Acquisition Device ID openrvdas
Data Acquisition Device Serial Port /dev/ttySONDE
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level primary
Enabled True
Vessel Installation Location
Vessel cms
Vessel Location Main Lab
Location Coordinates (meters)
Sensor Orientation vertical down
Height Above Waterline (meters) None
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Apr 02 2025 19:40:14     King, Tristan     sensor specs update     enabled changed from [False] to [True]    
Apr 02 2025 19:37:10     King, Tristan     new sensor     Copied sensor metadata from ysiexo102132 with modified serial number 23A104835.    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A:
  Feb 20 2025
Maintenance B:
 
Maintenance C:
 
Calibration
  2 months Nov 10 2024 Jan 10 2025
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
 
 
User Guides & Documents

Date Version Author Type Document Access
 
 
Ownership

Quantity Value
Owner other
Responsible Party other
Purchase Date
Acquire Date
Acquire Condition
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company
Website
Phone
Email
Shipping
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
OSU logo NSF logo RCRV logo
RCRV Datapresence and Engineering Support Center
Oregon State University
Corvallis, OR 97331, USA